Model
|
CMT-100
|
200,300
|
Max.Test Force(kN)
|
100
|
200,300
|
Accuracy Grade
|
0.5 Grade
|
Test force measurement range
|
0.2%~100%FS (Full scale)
|
Test force relative error value
|
Better ±0.5% than indicating value
|
Resolution of test force
|
±1/300000 of Max.test force(not marking graduation within full range, and same resolution all the way)
|
Deformation measurement range
|
0.2%~100%FS
|
Deformation and the relative error value
|
±0.5% of indicating value
|
Deformation resolution
|
1/300000 of max. deformation
|
Displacement indication relative error
|
Within ±0.5% of indicating value
|
Displacement resolution(μm)
|
0.025
|
Force control rate adjustment range
|
0.005~5%FS/s
|
Effective Tensile Space
(Customizable)
|
700mm
|
600mm
|
Effective Compression Space
(Customizable)
|
700mm
|
600mm
|
Effective Width Space
(Customizable)
|
550mm
|
Power supply
|
380V,50Hz,2~5kW
|
Working environment
|
Room temperature~35℃,relative humidity not more than 80%
|
Remarks
|
1. Equipped Fixture can be based on user’s needs or customer made
2. Alternative transducer is available to extend the measuring range
3. The tensile space can be heighten on user’s requirements |